Based on the Goldmann ...
... principle which is a world widely accepted gold standard, T170 applanation tonometer can provide fast and accurate IOP measurement data to clinical diagnosis and treatment.
Adjustable mechanical
... design ensures the reliable measurement result and durable performance of the instrument. Repeated measuring procedures do not reduce the natural ocular pressure of the patient.
In combination with
... all the tower type slit lamps. The measurement can be carried out on the patient sitting at the slit lamp in conjunction with other routine microscopic examinations.
Range of Measurement | 0 mmHg ~ 80mmHg(0 kPa ~ 10.64kPa) |
Moving Range of the Rings | 1.53x2=3.06mm |
Prism Diameter | 7mm |
Moving Range of Prism | 0~3mm |
Dimension | 190mmx80mmx80mm |
Net Weight | 480g |
Working Temperature | 0°C~40°C / 32°F~104°F |
Working Relative Humidity | 10%~85% |
Mount the Tonometer onto the Slit Lamp. Position the illumination of the Slit Lamp properly light up the Prism and switch the filter to the Cobalt Blue, in conjunction with the use of fluorescein, the 10P measurement value can be read from the dial drum of tomomter.